Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

Running average on the pressure sensor temperature #48

Open
TilmanJimenez opened this issue Nov 13, 2024 · 0 comments
Open

Running average on the pressure sensor temperature #48

TilmanJimenez opened this issue Nov 13, 2024 · 0 comments

Comments

@TilmanJimenez
Copy link

The SBE Data Processing Manual says:

Pressure temperature is computed using a backward-looking, 30-second running average, to prevent bit transitions in pressure temperature from causing small jumps in computed pressure. Because the heavily insulated pressure sensor has a thermal time constant on the order of one hour, the 30-second average does not significantly alter the computed pressure temperature.

but it seems that ctdproc doesn't do this. Is there a specific reason for this?

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment